BR-PV-IVT Impulse Voltage Test System
Implementation of standards: IEC 61730-2:2004 & Ed.2, VDE 0126-5:2008
The test equipment and procedure shall comply with IEC 60060-1. The values for T1 and T2 have been adjusted due to the variable and comparably high capacity of the many different samples a laboratory will perform the test on.
Maximum System Voltage | Impulse Voltage | |
Application class A | Application class B | |
1000 | 8000 | 6000 |
1500 | 10000 | 8000 |
Pulse voltage: 0.01~10KV (Max.12KV)
Loading waveform can be adjusted separately, Voltage comprehensive wave:
① IEC 61730 Ed.1& VDE 0126-5:2008: 1.2μs±30%/50μs±20%
② IEC 61730 Ed.2: 2μs±30%/20~50μs
Capacitance range of module: 6pF~250μF
(Crystalline silicon module about 300pF, thin-film module about 100μF)
Oscilloscope: including shielded boxes
脉冲电压发生器,BR-PV-IVT
脉冲电压发生器信息由上海泊睿科学仪器有限公司为您提供,如您想了解更多关于脉冲电压发生器报价、型号、参数等信息,欢迎来电或留言咨询。